HISTOPATHOLOGYà
HISTOLOGYà
AUTOMATIC
TISSUE PROCESSOR
|
|
SALIENT
FEATURES : |
|
| SECO-104
Automatic Tissue Processor-Electra Model (Micro Processor
Based) |
| Ø |
Under
the guidance of eminent pathologists our R & D Division
has developed yet another unique model of Tissue Processor
having complete Programmer logic Controller for complete
cycle. Timing & stage sequence, Wax bath temperature
safety system etc. are all controlled/operated by feather
touch key board with Digital Display.
|
| Ø |
The
entire unit looks very attractive with aesthetic appeal.
|
| Ø |
The
data is retained in case of power failure. Front panel
comprises of the key board, digital display, control switch.
The PID temperature controller have been used for controlling
the wax bath temperature. |
| TECHNICAL
DATA : |
| Ø |
The
12 stage timing sequence may be programmed for each stage
for duration between 1 minute to 9 hrs. 59 minute, in
steps of 1 minute. |
| DELAY
TIME : |
| Ø |
Delay
time can be programmed upto 99 hours and will be sequenced
with main control. |
| WAX
BATH SAFETY CONTROL : |
| Ø |
Wax
Bath PID temperature controller gives signal if temperature
is not between 50° to 65°C. Safety System will
function at 10th, 11th & 12th position and will stop
the main unit in stage 9 onward till the set temperature
is attained in the wax bath. |
| BASKET
ROTOR : |
| Ø |
Slowly
rotating 2 basket rotors are also operated through the
microprocessor.
The unit is supplied complete with 2 Nos. S.S. Basket
Rotors, 2 Nos. S.S. Baskets, 24 Nos. S.S. Tissue Capsules,
10 Nos. S.S. beakers without lip, 2 Nos. S.S. Wax Baths
(Special designed for use with PID temperature controller).
Suitable to work on 220 V, single phase, 50 Hz, AC supply. |
| Ø |
Capacity
:– 1 litre, 2 liters |
| Ø |
Spare
Accessories :– |
| |
- |
Automatic
Wax Bath specially designed for PID control and PT 100
sensor. It signal’s the machine in stage 9, prior
to immersion in Wax and if wax is melted it automatically
takes change from stage 9 to 10 and so on, but if wax
is not melted due to some reasons, then machine signals
“Wax not melted or temperature is not proper”.
When wax is melted/wax temperature is attained then only
the tissue basket is allowed to immerse or rotate in the
wax bath. Thus 100% safety of precious tissue. |
| |
- |
Basket
Rotor which rotate the tissue basket slowly. |
| |
- |
Tissue
Capsule Basket of S.S. with even perforation. |
| |
- |
Tissue
Capsule S.S. 28×28×9mm with sliding cover
|
| |
- |
Tissue
Capsule S.S. 20×20×10mm with sliding cover.
|
| |
- |
Tissue
capsule S.S. circular, 36mm dia with Auto press lid. |
| |
- |
Divider
for Tissue capsule ‘s’ shape. |
| |
- |
Divider
for Tissue capsule ‘v’ shape. |
| |
- |
Beaker
without lip. |
| |
|
|
|