X-RAY
GENERATOR |
| X-RAY
GENERATOR |
: |
High Frequency
40 KHz. X-Ray Generator (CE Certified). |
| Maximum Power |
: |
3.5 KW |
| Maximum mA output |
: |
100MA |
| Maximum KV output |
: |
35 KV |
| |
X-RAY
TUBE |
| X-Ray Tube |
: |
Rotating Anode X-Ray Tube |
| Focal Spot |
: |
- Small Focus = 0.1 mm
- Large Focus = 0.3 mm. |
| Target Material |
: |
Molybdenum. |
| |
CONTROL
PANEL |
| Control
Panel |
: |
Micro Processor controlled
Feather Touch Control Panel. |
| KV Range |
: |
22 to 35 KV in steps of 0.5
KV each. |
| mAs Range |
: |
1 mAs to 700 mAs |
| Technique selection |
: |
Manual Two Point Technique
(KV, mAs). |
| Digital Displays |
: |
- KV
- mAs |
| |
STAND
ASSEMBLY |
(A
compact Stand on which C-Arm containing X-Ray
Tube & Bucky Assembly is mounted) |
| Vertical Movement (Motor operated) |
: |
550 mm |
| Angular Movement of X-Ray
tube & Bucky |
: |
± 180° |
| Focus Film Distance |
: |
60 cm |
| Breast Compression |
: |
Pneumatically controlled with
Pneumatic Switch for activation & release. Adjustable
compression force. |
| Bucky |
: |
18 * 24 cm Bucky, Motor operated
Oscillating Grid of size 18 * 26 cm, 5:1, 30 lines/cm
focal distance 60 to 70 cm. |
| Collimator |
: |
Light Beam collimator with
Halogen Lamp with Auto shut off after 1 minute.
|
| |
POWER
SUPPLY REQUIREMENT |
| Power Supply |
: |
Single Phase, 230 Volts ±10%,
AC, 50 Hz, 15 Amps with Independent earthing on
the wall socket. |
| |
|
|