STERILIZATION
à
CYCLINDRICAL HORIZONTAL
STERILIZER |
|
SALIENT
FEATURES : |
|
| SECO-405
Steri H.P.Cylindrical Horizontal Sterilizer |
| Ø |
Introduction
The sterilizer is based on the principle that microbiological
action of saturated steam at elevated temperature is
rapid and through. All models work on the principle
of downward displacement of air which is the most economical
method of obtaining STERILIZATION
.
Robust and rigid construction, designed for all types
of bulk STERILIZATION
which are commonly needed in Medical,
Agricultural and Pharmaceutical Institutions.
Suitable for Sterilizing hospital dressings, linen,
rubber, plastic goods, surgical instruments, glasswares,
utensils etc. |
| Ø |
Construction
Triple walled with steam jacket and separate boiler
Inner chamber and steam jacket are made up of Heavy
gauge S.S. Sheet with leak proof argon-arc welding.
The sterilizer has single piece door made of stainless
steel. Back plate and ring is also made of thick stainless
steel sheet. All the sterilizers are hydraulically tested
to withstand 2.5 times the working pressure.
Mounted on tubular steel frame with ground levelling
screwed flanges. The Outer jacket is wrapped with asbestos
sheet or glass wool to minimise the heat losses due
to radiation and is covered by polished stainless steel
sheet for that elegant appearance.
|
| Ø |
Steam Generator
Made of heavy stainless steel sheet. Heavy ring mounted
in front of the boiler with folding thick stainless
steel
plate is fitted with heating elements and low water
level cutoff device to protect the former from burning
out dry.
Front folding plate system provides for easy cleaning
of the deposited scale on the elements and the walls
for long life and efficiency. Fitted with water gauge
glass for water level indication, water inlet and outlet
valves.
|
| Ø |
Pressure
Control
The PIEZOSTAT (Pressure control device)
is incorporated in all electrically operated sterilizers.
It economises on power consumption and reduces the frequent
opening of the steam release valve and thus prevents
the release of steam in the room, It cuts off the power
supply to heaters when the set pressure is achieved
and re-energises the heating elements when pressure
falls below the set point. |
| Ø |
Heating Elements
Flanged type, immersion heating elements are made of
high grade material and duly ISI marked.
|
| Ø |
Safety Door
Sterilizer has hinged self-locking single piece thick
S.S. plate door and cannot be opened when under pressure.
The self-locking device automatically disengages the thread
mechanism when the sterilizer is under pressure to preclude
any eventuality of accident. |
| Ø |
Single Point Control
Complete sterilizing cycle is controlled from one point
with the help of Multiport Valve fitted
at front top. Two speeds of steam exhaust are available;
fast and slow.
Easily readable jacket and chamber pressure on gauges
mounted on the multi port valve. A moisture trap is
fitted in the chamber discharge line to absorb the condensate
automatically to prevent moistening of the subject matter
to be sterilized.
|
| Ø |
Switch Box
Houses the PIEZOSTAT, main switch, pilot lamps, relays
etc. and mounted on the stand of the sterilizer for
easy operation.
|
| |
- |
Other Features
|
|
| |
- |
TRIPLE
SAFETY : The Sterilizer is provided with
the triple safety features. At the boiler level by a PIEZOSTAT
which automatically limits the pressure to the required
set value and a spring loaded safety valve in case of
its failure which releases the steam to keep pressure
within the safety limits. At the chamber level, a spring
loaded safety valve and a dead weight release valve to
release the steam in case the pressure exceeds the safety
limit. All these safety features function independent
of one another and sequentially i.e. one takes over the
charge in case of others failure. |
| |
- |
A plug screen prevents
the discharge line from choking. The plug is easily removable
for daily cleaning. |
| |
- |
Powerful Ejector for
drying sterilized linen circulates air throughout the
chamber. The circulating air passes through a corrosion
resistant metallic wool filter. |
| |
- |
A dial thermometer is
provided to show the chamber temperature. |
| |
- |
Automatic Vacuum breaker
is provided to break vacuum in case of formation of vacuum
due to steam condensation. |
| |
- |
Operating Temp. &
Pressure. |
| |
- |
Sterilizing Temp : 120°C |
| |
|
Sterilizing pressure
: 1.2 to 1.5 kg/cm2 (15 psi to 22psi) |
| |
- |
Power Requirement :
Suitable to operate on 440V volts, 3 ph, 50 Hz, Ac supply |
| |
- |
Available in following
capacities |
| |
- |
dia × depth load |
| |
- |
400 × 600 mm 6
kw |
|
| |
- |
400 × 1100 mm
9 kw |
|
| |
- |
500 × 900 mm 9
kw |
|
| |
- |
500 × 1200 mm
9 kw |
|
| |
- |
Optional Accessories
|
|
| |
- |
Chamber/Jacket of S.S.
316 |
|
| |
- |
Door of S.S. 304/316 |
|
| |
- |
Dressing Drums |
|
| |
- |
Direct steam supply
from centralised boiler |
|
| |
- |
Temperature recorder |
|
| |
- |
High pressure High vacuum
model |
|
| |
- |
Additional door (Double
door) model. |
|
| |
- |
Fully automatic micro
processor controlled model |
| |
- |
Spare Accessories |
|
| |
- |
Pressure gauge/Compound
gauge |
|
| |
- |
Water level gauge glass |
|
| |
- |
Dial thermometer
|
|
| |
- |
Non return valve |
|
| |
- |
Moisture trap |
|
| |
- |
Jointless Gasket :
Neoprene/silicon rubber |
| |
- |
size 400mm & 500mm |
|
| |
- |
Heating elements (Flange
Type) Rating kw 6.0, 9.0 |