STERILIZATION à
HIGH PRESSURE HIGH SPEED HORIZONTAL CYLINDRICAL |
|
SALIENT
FEATURES : |
|
| SECO-611
High Pressure High Speed Horizontal Cylindrical/ Rectangular
Regulated Medical Wast lizer |
| Ø |
Purpose :
Equipment is designed for Sterilization of regulated biomedical
waste generated from hospital for destruction of bacteria
and related organism (Spores).
|
| Ø |
Design
: The lizer equipment to treat medical
waste is designed for following parameters. |
| |
- |
For gravity flow autoclave
|
|
| |
- |
The temperature of not
less than 121°C and pressure of not less than 1.02
K.G./CM2 and autoclave residence time not less than 60
minutes. |
| |
- |
The temperature of not
less 135°C and pressure of not less than 2.1 K.G./CM2
and autoclave residence time not less than 45 minute.
|
| |
- |
The temperature of not
less than 149°C and pressure of not less than 3.5
K.G./CM2 and autoclave residence time not less than 30
minute. |
| |
- |
For vacuum autoclaving
with pre vac & purge system. |
| |
- |
The temperature of not
less than 121°C and pressure of not less than 1.02
K.G./CM2 and autoclave residence time not less than 45
minutes with one or more pre vacuum pulse. |
| |
- |
The temperature of not
less than 135°C and pressure of not less than 2.1
K.G./CM2 and autoclave residence time not less than 30
minutes with one or more pre vacuum pulse. |
| |
Construction :
|
| |
- |
Shape
: Either cylindrical/rectangular. Chamber
is made of boiler grade sheet/AISI 304/AISI 316 thick
sheet having one or two nos. pivoted door mounted on a
sturdy structural stand dully insulated & covered
either by S.S. or M.S. Sheet. Pit mounted design is also
available as per the customer requirement for easy loading
& unloading. The door will be provided with automatic
safety door looking / unlocking mechanism. Valves are
provided for operation safety steam inlet, drainage etc.
|
| |
- |
Control and Monitoring
System : |
|
| |
- |
Microprocessor based
computer system and printer is provided for controlling
and monitoring the lizing cycle with the following feature |
| |
- |
Fully programmable microprocessor
based unit with no volatile memory and with different
program combinations. |
| |
- |
Automatic operation
of the lizer. |
|
| |
- |
Self diagnostic system
gives fault massage to the operator. |
| |
- |
Manual operation facility
for stand by program in case of control failure |
| |
- |
Cycle progress and information
indicating any error massage is displayed and recorded
on the printer. The printer shall also record data time,
day load identification no. and operating parameters such
as temperature of the steam in chamber, steam pressure,
vacuum time and temperature of steam condensate after
cooling through out the entire length of autoclave cycle.
|
| |
- |
Available
in following Sizes : |
| |
|
Cylindrical
Model : |
|
| |
- |
400 (Dia) × 600
mm (Depth) |
|
| |
- |
500 (Dia) × 1200
mm (Depth) |
|
| |
|
Rectangular
Model : |
|
| |
- |
600 (H) × 600
(W) × 1200 mm (D) |
| |
- |
900 (H) × 600
(W) × 1500 mm (D) |
| |
- |
Other custom built sizes
are also available. |
|